ADI ARTECH

60210

Single Point Platform Load Cell

Description

Application

Key Features

Technical Specifications

STANDARD CAPACITY

15, 22.5, 45, 90  -Kgf.

1,1.5,2,2.5,3,5,7.5,10,15,20, 30          – Tf.

PERFORMANCE

PERFORMANCE

Non- Linearity

Hysteresis

Non-Repeatability

Creep(30minutes)

ZeroBalance

< ±0.025% FSO (Full Scale Output)

< ±0.01% FSO

< ±0.02% FSO

< ±0.03% FSO

 ±1.0% FSO

ELECTRICAL

ELECTRICAL

Output

Excitation Voltage
Input Resistance
Output Resistance

Insulation Resistance

Maximum 15 VDC (Suggested 5VDc / 10VDc)

390 ± 15.0 Ω (Ohms)

350 ± 5.0 Ω (Ohms)

(100 VDC) > 1000 Mega Ω (Ohms)

2.00 mV/V ± 0.2%

STRUCTURE

STRUCTURE

Safe Overload Of Rated Capacity
Ultimate Overload of Rated Capacity
Allowable Side Load
Deflection

Platter Size

Finish Product

As per IS/IEC 60529:2001

150%
250%
50%
< 0.4 mm at FSO

500 mm X 500 mm

Electroless Nickel Plated Tool Steel

Meets IP 65

TEMP.

Operating Temperature

Compensated Temperature
Temperature Effect Output

Temperature Effect Zero

0ºC to 60ºC

10ºC to 50ºC

< 0.0025 % FSO/ ºC

< 0.0030 % FSO/ ºC

OPTIONS

Extra cable length
Digital Version

Compatible Instrument

Specify extra length In Meters

Digital Output

DLC-6M, DLC-6, Summing Junction Box

Industries

Weighing